Model: M10
◎Microprocessor based servo controlled temperature system
◎Control mode: Air mode
◎Set temperature, air temperature, heating power,timer are displayed separately by LED
◎Self - testing function, various failure alarms by audible and visual
◎>37℃ temperature set function
◎Triple protection for over temperature with individual cut off device, more safety system
◎ The inclination of infant bed is adjustable
◎Humidity is adjustable continuously
◎Low noise fan
◎Single wall hood, 4 operating windows ,2 iris ports and 4 infusion seals
Specification
Power Supply |
AC110/220v 50~60HZ |
Power input |
500VA |
Air Temp. control range |
25.0℃~ 37.0℃;37.1℃~ 39.0℃ |
Air Temp. Display range |
20.0℃~ 42.0℃ |
Air temp.sensor accuracy |
±0.5℃ |
Air temp.deviation alarm |
±3.0℃ |
Uniformity of mattress temp. |
≤0.8℃
|
Warm up time |
≤45min
|
Noise level |
≤50dB(A)
|
Mattress size |
370*630*20mm
|
Water box capacity |
1200ml |
Operating conditions |
Environment temperature 20℃~30℃ |
Environment relative humidity 30%~75%RH |
|
Atmospheric Pressure 700~1060hPa |
|
Environment air velocity of flow <0.3m/s |
Optional function:
Up LED phototherapy unit H60
Down phototherapy unit H80
X-ray cassette holder
Height adjustable
LCD touch screen
Transport and Storage |
Environment Temperature:-10℃~ +55℃ |
Environment Relative Humidity: ≤95%RH |
Atmospheric Pressure: 500~1060hPa |